Martin:
Attached is complete set of BB and SP instrument calibration data. Below is instrument configuration information, including NC27 and NCHR substitutions.
Bob
BB = broadband SP = short period SMA = Strong Motion Accelerometer
Here is manufacturer's calibration data for
BB seismometers (Guralp)
SMA (Guralp)
SP seismometers (Geosense: sensor) (Guralp: digitizer)
Attached are...
Four CMG-1T BB seismometer sensor calibration sheets: T1063, T1064, T1065,T1067
Four CMG-DM24 BB digitiser calibration sheets: A1716, A1717, A1719, A2260
Three CMG-5T SMA sheets: T5O75, T5O76, T5O77
One Geosense BH-1 SP seismometer sensor specification sheet (generic)
One BH-1-15 SP seismometer calibration sheet (consider generic)
One Guralp CMG-DM24 SP digitizer calibration sheet for s/n A2179, A2042, A2180, A2181
BB Instrument configurations in 2009 were as follows...
NCBC T1064, T5O76, A1719
NC89 T1063, T5O75, A1717
NC27 T1067, T5O77, A1716
BB Instrument configuration change from Oct.6, 2010...
NC27 T1065, T5S27, A2260
SP Instrument configurations in 2010 as follows...
KEMF BH-1-17 A1349
NCHR BH-1-16 A1350
SP Instrument configuration change from Sep.14, 2011
NCHR BH-1-18 A1352
Site History 2013-08-01
################################################################################# NC27 ==== Station Installation - September 14, 2009 ========================================= Seismometer @ 100sps, 2sps (ttl 24) Accelerometer @ 100sps, 2sps (ttl 24) Pressure Sensor @ 2sps (ttl 24) October 5, 2010 23:59:59 ======================== Seismometer, accelerometer and pressure sensor deactivated June 10, 2010 17:43:00 ====================== Seismometer @ 100sps, 1sps (ttl 49) Seismometer Mass Positions @ 1sps Accelerometer @ 100sps, 1sps (ttl 49) Pressure Sensor @ 1sps (ttl 49) July 17, 2013 17:55:00 -- Seismometer seems to have problems with these settings stilll trouble shooting ====================== Upgraded DM24 firmware version 50xxxx from 40yyy Seismometer @ 200sps, 40sps, 8sps (ttl 38) Seismometer Mass Positions @ 1sps Accelerometer @ 200sps, 40sps, 8sps (ttl 38) Pressure Sensor @ 8sps (ttl 38) July 19, 2013 17:20:00 ====================== Datalogger DM24 not stable at new firmware version so reinstalled previous firmware version. Therefore had to revert to previous sample rates. Seismometer @ 100sps, 1sps (ttl 49) Seismometer Mass Positions @ 1sps Accelerometer @ 100sps, 1sps (ttl 49) Pressure Sensor @ 1sps (ttl 49) ################################################################################# NC89 ==== Station Installation - September 17, 2009 ========================================= Seismometer @ 100sps, 2sps (ttl 24) Accelerometer @ 100sps, 2sps (ttl 24) Pressure Sensor @ 2sps (ttl 24) September 10, 2010 22:50:15 =========================== Seismometer, accelerometer and pressure sensor deactivated September 10, 2010 22:51:30 =========================== Seismometer @ 100sps, 1sps (ttl 49) Accelerometer @ 100sps, 1sps (ttl 49) Pressure Sensore @ 1sps (ttl 49) March 1, 2013 22:30:59 ====================== Seismometer, accelerometer and pressure sensor deactivated March 1, 2013 22:31:00 ====================== Seismometer @ 200sps, 40sps, 8sps (ttl 38) Accelerometer @ 200sps, 40sps, 8sps (ttl 38) Pressure Sensor @ 8sps (ttl 38) ################################################################################# NCBC ==== Station Installation - September 07, 2009 ========================================= Seismometer @ 100sps, 2sps (ttl 24) Accelerometer @ 100sps, 2sps (ttl 24) September 11, 2010 01:14:00 =========================== Seismometer and accelerometer deactivated September 11, 2010 01:15:00 =========================== Seismometer @ 100sps, 1sps (ttl 49) Accelerometer @ 100sps, 1sps (ttl 49) March 1, 2013 20:00:00 ====================== Seismometer @ 200sps, 40sps, 8sps (ttl 38) Accelerometer @ 200sps, 40sps, 8sps (ttl 38) ################################################################################# KEMF ==== Station Installation - September 30, 2010 ========================================= Seismometer @ 100sps, 1sps (ttl 49) July 15, 2013 19:00:00 ====================== Seismometer @ 200sps, 40sps, 8 sps (ttl 38) ################################################################################# NCHR ==== Station Installation - October 04, 2010 ======================================= Seismometer @ 100sps, 1sps (ttl 49) September 13, 2011 23:59:59 =========================== Seismometer deactivated September 14, 2011 18:48:50 =========================== new seismometer & datalogger (see dbbuild file for s/n's) Seismometer @ 100sps, 1sps (ttl 49) #################################################################################